Mandatory Regulations for Designing Dental X-Ray Rooms According to the Latest Ministry of Health Standards
The dental X-ray room is a specialized functional area subject to strict regulation regarding radiation safety, equipment licensing, and shielding standards. If the design does not meet regulatory requirements, the clinic will not be permitted to operate the X-ray equipment, even if the machine has already been installed.
This article provides a comprehensive summary of legal regulations, technical standards, and practical inspection experience in accordance with the latest guidelines issued by the Ministry of Health and relevant specialized regulatory authorities.

1. Legal Basis for Designing a Dental X-Ray Room
The design and operation of a dental X-ray room must comply with:
1.1 Laws and Regulations on Medical Examination and Treatment
Law on Medical Examination and Treatment No. 15/2023/QH15
Regulations on facility conditions and safety when using medical equipment.
1.2 Radiation Safety Regulations
-
A license for the use of radiation equipment must be obtained before operation.
-
Radiation safety inspection documentation is required.
-
A trained radiation safety officer must be appointed.
1.3 Licensing Authority
Inspection and licensing are conducted by the Department of Health in coordination with the radiation regulatory authority.
Failure to meet legal requirements → X-ray imaging is not permitted.
2. Standard Area for a Dental X-Ray Room
Minimum area: 6 – 8 m²
The space must accommodate:
-
X-ray machine
-
Patient chair
-
Radiation safety distance
The X-ray room must not be placed within the treatment area unless compliant radiation shielding is installed.
A small room may still be licensed if radiation shielding meets standards.
3. Mandatory X-Ray Shielding Requirements (Key Licensing Factor)
This is the most critical criterion during inspection.
3.1 X-Ray Room Walls
Radiation shielding must be ensured using one of the following solutions:
-
Solid brick wall ≥ 20 cm thick
-
Solid concrete meeting radiation standards
-
Specialized lead shielding panels
If radiation leakage is detected → operation license will not be granted.
3.2 X-Ray Room Door
-
Lead-lined radiation-shielded door is mandatory
-
Must prevent radiation leakage completely
3.3 Observation Window
-
Lead glass radiation shielding is mandatory
-
Installed at the control area
4. X-Ray Control Area (Mandatory)
A safe control position must be provided:
-
Located outside the direct radiation exposure zone
-
Properly shielded from radiation
-
Equipped with lead glass for patient observation
-
Remote operation control
Standing next to the machine during exposure is not permitted.
5. Radiation Safety Standards in the X-Ray Room
5.1 Warning Signs
Mandatory:
-
Radiation warning sign
-
Indicator light when X-rays are being emitted
5.2 Protective Equipment
Must include:
-
Lead protective apron
-
Thyroid lead collar
-
Lead shielding screen (if necessary)
5.3 Exposure Control
-
No unauthorized persons allowed in the room
-
Avoid unnecessary radiation exposure
6. Radiation Safety Personnel Requirements
The clinic must appoint a person responsible for:
-
Radiation safety training
-
Understanding machine operation procedures
-
Handling radiation incidents
Without this personnel → licensing approval will not be granted.
7. Regulations on Dental X-Ray Equipment
The equipment must:
-
Have clear origin documentation
-
Undergo radiation inspection
-
Have complete technical documentation
-
Be officially registered for use
Uninspected equipment → operation is not permitted.
8. Documentation for Applying for X-Ray Room Authorization
Includes:
-
X-ray equipment documentation
-
Radiation shielding design drawings
-
Radiation safety documentation
-
Radiation safety training certificate
-
Equipment inspection certificate
-
Radiation protection documentation
9. Common Design Errors Leading to License Rejection
-
No radiation shielding for walls – door – glass
-
No safe control area
-
No radiation warning signage
-
No protective lead apron
-
No equipment inspection
-
No appointed radiation safety officer
These are common mistakes when self-designing.
10. Practical Experience for Fast X-Ray Licensing Approval
-
Design radiation shielding from the beginning
-
Use certified standard lead materials
-
Arrange the control area correctly
-
Prepare complete radiation safety documentation
-
Inspect equipment before submitting the application
Proper design helps:
-
Avoid application rejection
-
Avoid costly renovation
-
Obtain licensing approval quickly
-
Ensure long-term safe operation
11. The Role of X-Ray Room Design in Dental Clinic Licensing
In practice, more than 50% of clinics experience licensing delays due to incorrect X-ray room design. Proper design not only legalizes equipment use but also ensures:
-
Safety for dentists and patients
-
Compliance with Ministry of Health standards
-
Long-term risk-free operation
-
Enhanced clinic credibility
12. Frequently Asked Questions About Dental X-Ray Rooms
12.1 Is X-ray shielding mandatory?
Yes. Without shielding → licensing approval will not be granted.
12.2 Can a small room be used for X-ray?
Yes, if proper shielding and radiation safety standards are met.
12.3 Is a radiation safety officer mandatory?
Yes. This is a compulsory requirement.
12.4 How long does it take to obtain X-ray licensing approval?
Typically 30 – 45 days if the documentation meets standards.
13. Conclusion
Designing a dental X-ray room must strictly comply with radiation shielding standards, radiation safety regulations, and legal documentation requirements. Standardization from the beginning is the decisive factor for fast licensing approval, safe operation, and full compliance with Ministry of Health regulations.
Contact for Medical-Standard Dental Clinic Design Consultation
-
Website: https://thietkenhakhoa.com/
-
Hotline: 02866525368 – 0978244411 – 0935650368
-
HCMC Office: No. 40, Street 176, Phuoc Long A, Thu Đức City.
-
Da Nang Office: K29/150 Tran Ngoc Suong, Hoa Tho Dong, Cam Le District.
14. INTERNAL LINK MAP (By Functional Area)
Dental Facade: [Clinic Frontage Design] | [Signage Construction] | [Brand Identity]
Reception: [Reception Desk Design] | [Waiting Lounge Interior]
Treatment Room: [Medical-Standard Design] | [Ergonomic Chair Layout] | [MEP Systems]
Implant Suite: [Implant Room Design] | [Sterile Standards] | [Central Vacuum Systems]
X-Ray Room: [Radiology Room Design] | [Lead Shielding Standards]
Sterilization: [Sterilization Room Design] | [Infection Control Workflow] | [Autoclave Integration]
Consultation: [Consultation Room Design] | [Patient Experience Design]
Staff Area: [Staff Workspace] | [Doctor’s Office Design]
Restrooms: [Clinic Toilet Design] | [Hygiene Standards] | [Advanced Plumbing]